ECG Leads | 10 lead 12 channel simulataneous 10-second resting ECG |
Dimension | 296 (w) x 305.5 (h) x 92.5 (d) mm |
| Approximately 2.98 kg |
Sampling rate | Analysis : 500 / sec’ |
| Digital : 8,000 / sec’ |
Recording Channel | 3, 6, 12 channel and 1 channel (60 sec) |
Sensitivity | 5, 10, 20 auto (I~aVF: 10, V1~V6:5) mm/mV |
Printing Speed | 12.5, 25, 50 mm/s |
Filters | AC (50/60 Hz, -20dB or better) |
| Muscle (25~35Hz, -3dB or better) |
| Base Line Drift (0.1Hz, -3dB or better) |
| Low pass filter (off, 40Hz, 100Hz, 150Hz) |
Display | 2 x 16 character LCD display |
LCD Display | Sensitivity, speed, filter status, HR, printing form, rhythm lead |
LED Indicator | Signal quality, power source, battery |
Keyboard | Membrane keyboard |
| Alphanumeric and symbol available |
Patient Data | ID, name, age, sex, height, weight |
Basic Measurement | Heart rate, PR, QRS, QT/QTc, P-R-T axis |
Recorder | Thermal print head, thermal roll paper |
| Report papers: |
| -Width: A4: 210mm or 8.5" |
| -Length: A4: 300mm or 11" |
| Effective writing width: 204mm |
| Resolution: |
| - Vertical: 8 dots/mm |
| - Horizontal: 16 dots/mm |
Electrical | Internal noise: 20uV(p-p) max |
| Input circuit: Floating input |
| Isolated and defibrillation protected |
| Input impedance: >10MQ |
| Input range: > +/- 5mV |
| Common Mode Rejection: > +100dB |
| DC offset voltage: > +/- 300mV |
| Time constant: 3.2 sec |
| Patient leakage current: |
| Frequency response: 0.05~150Hz |
Signal Quality Control | Disconnected lead detection |
Power | Power Supply: AC or built-in Battery (option) |
| Power requirement: 100 - 240 Vac, 50/60Hz, 1.0 - 0.5A |
| Power consumption: 60w max |
Battery Capacity | 6 hours of normal use (approximately 100 automatic ECG printouts) |
Communication | PC connection with LAN |
Safety | Class 1, Type BF |
Conformity | CE, CSA |
Environment | Operating humidity: 30~85% |
| Operating temperature: 10° ~40° C |
| Atmospheric pressure: 70 ~ 106 KPa |