Cardio 7


Capture, store, analyze and transmit with streamlining workflow

 

Features
Simple Fast acquisition
  • One button “AUTO” operation from acquisition to print-out & data record in 3 sec’.
Easy ECG file data conversion
  • Cardio 7 supports easy file conversion of PDF, XML, JPG, BMP, MFER, DICOM to transfer with easy user interface.
Enhanced sampling rate for data accuracy
  • With 8,000Hz per second sampling rate, Cardio 7 is designed to improve clinical decision for cardiology with accurate ECG interpretation algorithm as well as enhanced 200 ECG / 5min’ data storage.
Advanced algorithm & Analysis
  • With long term recording of 3, 5, 10min’, Cardio Touch has various analysis algorithms of ECG diagnosis.
Patient data management in mind
  • All patient data is transferrable to BMS Plus, PC based data management SW with real time remote operation.
BMS plus offers caregivers the workflow flexibility for
  • Data preview
  • Device remote control
  • File conversion to JPG, XML, DICOM, PDF, MFER for data transfer thru PACS
  • Direct print-out, magnifying ECG trace, measure ECG complex, adding comments, ECG trace comparison

User-friendly interface & diagnosis

  • Full touch operation with intuitive 6 buttons for data acquisition, interpretation, recording and file transfer to EMR
  • Quick guide indication at printer cover for easy operation recognition
  • Thermal A4 paper compatible with grid printing mode
  • 120 types of interpretation algorithm integrated
  • Text, Guid, Vector print-out format for enhanced interpretation
  • Rechargeable Li-ion battery for 3hr operation / 200 pages printout
  • Long term recording 3,5 up to 10min’

Spirometry connection ( Option )

  • Support various pulmonary function test : FVC, SVC, MVV
  • Combined patient data storage with ECG acquisition
  • Various formular integration : Morris-Polgar, ECCG-Quanher, Knudson-ITS, Pereira, KOREA CJK
  • Bionet’s unique infection free disposable mouthpiece solution

Network & Data management

  • Wired / Wireless network connectivity to HIS / EMR with easy DHCP interface
  • Various file conversion to JPEG, BMP, PDF, XML, MFER and DICOM
  • Direct DICOM file transfer to PACS
  • 200 internal memory
  • Full disclosure up to 5min’ for HRV, Arrhythmia analysis
  • BMS PLUS ( Bionet ECG SW ) connection for enhanced data management to PC thru LAN for data transfer to HIS ( PACS/EMR ) with various file format conversions of JPEG, BMP, XML, PDF, MFER, DICOM.

Bionet Cardio Series with New Logo

Bionet Cardio7 Video Manual

How to use Bionet BMS Plus

How to connect BMS Plus to Bionet ECG machine

How to install Bionet BMS Plus

Bionet Cardio7 Barcode ID scan

wdt_ID Item Type Download
1 blank Brochure Download
3 blank Download Download
Product SpecificationsValue
ECG LeadsSimultaneous 12 chnnel resting ECG
Recording Channel3CH+1RHY, 3CH+3RHY, 6CH+1RHY, 12CH,
1CH Long Time(1min, 3min, 5min,10min) and Beat Report
Gain2.5, 5, 10, 20, Auto (I~aVF: 10, V1~V6: 5) mm/mV
Printing Speed5, 12.5, 25, 50, 100 mm/sec
Sampling RateAnalysis Sampling Rate - 500
Digital Sampling Rate - 8,000
FiltersAC (50/60 Hz, -20dB or better),
Muscle (25~35Hz, -3dB or better),
Base line drift (0.05Hz, 0.1Hz, 0.2Hz, -3dB or better),
Low pass filter(off, 40Hz, 100Hz, 150Hz)
Display800 x 480 pixel resolution, 7” Color TFT wide display
12 channels preview, ID, HR, Gain, Speed, Power status
User InterfaceTouch screen, Rotary push-knob
Printer ResolutionThermal print head, Roll Paper
Report paper
- width : A4/Letter 215mm(8.5”)
- length : A4 297 mm (11.7”)
Letter 279mm(11”)
- Resolution: 8dot/mm(0.125mm pitch)
Patient dataID, name, age, gender, height, weight, smoke,
department, room number, race
Basic MeasurementHeart rate(30~300bpm ±2), PR, QRS, QT/QTc, P-R-T axis,
SV1/RV5/R+S amp.
ECG Data StorageInternal Storage for 200 ECGs : Built- in memory
ElectricalInternal noise: 20uV(p-p)max
Input impedance: More than 50MΩ
Input voltage range: ±5mV
CMRR: Greater than 105dB
DC offset voltage: ≥ ±400mV
Patient leakage current: < 10uA
Frequency response: 0.05 ~ 150 with in –3dB
Isolated , defibrillation and ESU protected
Signal QualityPacemaker pulse detection
ControlLead fault detection , Signal Saturation detection
Line PowerInput:100 - 240 V~ 2-1A, 50/60Hz, Output : 15VDC, 4A 60VA
Battery typeReplaceable and rechargeable, Lithium ion , 11.1V , 2600mA
Battery Capacity360 minutes of continuous operation without recording or
200 ECGs in 12 channel format at 25mm/s and 10mm/mV
with a battery charged during approximately 3 hours from total discharge(with display off)
CommunicationLAN, WIFI(option), USB flash memory, USB barcode scanner
Safety ConformityClass I, Type CF defibrillation-proof applied part.
EnvironmentalOperating temperature: 5 ~ 40℃
Operating humidity: 10~90%RH
Atmospheric pressure: 70 ~ 106KPa
Dimensions300(W) x 299(H) x 123(D) mm , Approx. 3.5 kg(Main Body)
Standard AccessoryPatient cable (1 EA), Limbs electrodes (1 SET),
Chest electrodes(1 SET), ECG chart paper(1 EA),
Power cord (1 EA),
Operation manual & ECG diagnosis guide (1EA)
OptionsRechargeable battery(1 EA),
Cart(1 EA), Hanger(1 EA)
blank